Le. Leon et J. Catapano, INDIRECT FLOW-INJECTION ANALYSIS OF ASCORBIC-ACID BY PHOTOCHEMICAL REDUCTION OF METHYLENE-BLUE, Analytical letters, 26(8), 1993, pp. 1741-1750
Ascorbic acid was determined indirectly by spectrophotometry at 666 nm
based on the photochemical reduction of methylene blue using flow-inj
ection analysis. The carrier stream was 12.7 mugml-1 methylene blue in
phthalate-HCl buffer solution at pH 3.2. The reactor was irradiated w
ith a 500-Watt halogen lamp to facilitate the development of the photo
chemical reaction. The system allows determination of ascorbic acid in
the range 0.18-6.12 mugml-1 with relative standard deviations of 2.09
and 0.31% for 1.97 and 4.92 mugml-1 samples, respectively, and a samp
ling frequency of 50-55 h-1. The proposed method was applied successfu
lly to the determination of ascorbic acid in vitamin C tablets.