A. Paquignon et al., EVALUATION OF THE CLINITEK-200 URINARY TEST-STRIP READER IN THE ANALYSIS OF DOG AND RAT URINES IN PRECLINICAL TOXICOLOGY STUDIES, Laboratory animals, 27(3), 1993, pp. 240-246
The performance of a clinical urinary test-strip reader Clinitek 200 w
as evaluated for dog and rat urines, in the context of pre-clinical to
xicology studies. No major discrepancies were found between data gener
ated by visual estimation or automatic measurement. Analysis of spiked
samples showed good agreement between actual concentrations and Clini
tek 200 responses for ketone bodies and glucose although a lack of sen
sitivity was found for the latter. Results for proteins showed over- o
r underestimation in dog and rat urines respectively at low concentrat
ions, and overestimation at high concentrations in both species. Repro
ducibility of responses was excellent for ketone bodies, glucose and p
roteins but was weaker for haemoglobin and bilirubin. High bilirubin c
oncentrations were found to interfere with the haemoglobin reaction. T
he pH measurements were found to be accurate only around pH 7. Specifi
c gravity measurements were unreliable. Overall, the Clinitek 200 as a
screening tool proved sufficiently reliable in the measurement of all
parameters tested, with the exception of specific gravity.