SELECTIVE BIOCATALYTIC MEASURING CELL FOR FLOW-INJECTION ANALYSIS

Citation
J. Tarcali et al., SELECTIVE BIOCATALYTIC MEASURING CELL FOR FLOW-INJECTION ANALYSIS, Electroanalysis, 5(4), 1993, pp. 277-281
Citations number
10
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
10400397
Volume
5
Issue
4
Year of publication
1993
Pages
277 - 281
Database
ISI
SICI code
1040-0397(1993)5:4<277:SBMCFF>2.0.ZU;2-C
Abstract
A flow-through measuring cell developed for flow injection analysis (F IA) measurements is described. The cell works detecting amperometrical ly the hydrogen peroxide formed in an enzyme-catalyzed oxidation react ion of different substrates. he electrode consists of a thin, hollow d ialysis fiber introduced into the flow analysis channel, an enzyme sol ution film pumped inside the fiber, and a cellulose acetate protective layer-coated thin platinum wire-base sensor housed inside the hollow fiber. The measuring cell for two different model compounds, glucose a nd ethanol, was well applicable for the concentration determinations i n small volume solution aliquots. Important advantages of the cell are its versatility and its easy activation for measurements.