Objective. The purpose of this study was to determine the accuracy of
five rotating vane or turbine-type respirometers-fdE Magtrak Respirato
ry Monitor, fdE Haloscale Standard Wright Respirometer, Micro-Medical
Pocket Monitor II, and Ohmeda 5400 and 5410 Volume Monitors-which are
marketed as suitable for measuring expired gas volume in anesthetized
adult patients. Methods. The devices were tested for accuracy at simul
ated minute ventilations ranging from 2.4 to 15 L/min with three gas m
ixtures: air, air saturated with water vapor, and N2O/O2 (70:30). Resu
lts. In the usual range for adult minute ventilation during general an
esthesia with intermittent positive-pressure ventilation (5-12 L/min),
all devices were on average accurate to within 12%. At the lowest min
ute volume readings (2.4 L/min) all devices except the Pocket Monitor
tended to have low readings. No significant differences in readings we
re introduced by either saturation with water vapor or introduction of
70% N2O, except with ventilation below 4 L/min with both Ohmeda respi
rometers. Conclusions. All monitors showed reasonable clinical accurac
y in the usual adult minute ventilation range during general anesthesi
a with intermittent positive-pressure ventilation (> 5 L/min) when use
d to measure air, water vapor-saturated air, or N2O/O2 mixtures.