COMBINED TSL-ESR MGO DETECTORS FOR IONIZING AND UV RADIATIONS

Citation
Vs. Kortov et al., COMBINED TSL-ESR MGO DETECTORS FOR IONIZING AND UV RADIATIONS, Radiation protection dosimetry, 47(1-4), 1993, pp. 273-276
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
47
Issue
1-4
Year of publication
1993
Pages
273 - 276
Database
ISI
SICI code
0144-8420(1993)47:1-4<273:CTMDFI>2.0.ZU;2-N
Abstract
MgO single crystals doped with iron after a reducing thermochemical tr eatment can be used for measuring ionising and UV radiation doses with simultaneous readout of TSL and ESR. A single maximum at 400 K is obs erved in TSL spectra. Its intensity grows in a linear fashion with the radiation dose from 10(-4) to 10(-4) Gy for X radiation and from 10(- 1) to 10(3) muJ.cm-2 doe UV radiation (lambda = 250 nm). Dose dependen ce of the ESR signal intensity caused by Fe3+ ions is linear from 1 to 10(4) Gy for X irradiation. Under UV irradiation the ESR signal is re gistered starting from doses of 2 x 10(2) muJ.cm-2. The upper level of the measured dose exceeds 10(5) muJ.cm-2. Complete anneal of the ESR signal is observed at 950 K. This makes it possible to use the detecto rs when measuring irradiation doses at higher temperatures.