SPLIT SHIELD FOR CORRECTION OF RADIATION ERRORS IN STEADY-STATE THERMAL-CONDUCTIVITY MEASUREMENTS

Citation
J. Talpe et A. Avendano, SPLIT SHIELD FOR CORRECTION OF RADIATION ERRORS IN STEADY-STATE THERMAL-CONDUCTIVITY MEASUREMENTS, Cryogenics, 33(8), 1993, pp. 833-835
Citations number
6
Categorie Soggetti
Physics, Applied
Journal title
ISSN journal
00112275
Volume
33
Issue
8
Year of publication
1993
Pages
833 - 835
Database
ISI
SICI code
0011-2275(1993)33:8<833:SSFCOR>2.0.ZU;2-6
Abstract
An axial heat flow, steady state method for thermal conductivity measu rements above 80 K on poor conductors is described. A split shield (on e part held at the sink temperature and the other kept at the hot-end temperature) is shown to be successful for the correction of radiation errors. Measurements on poly(methyl methacrylate) (PMMA) are reported .