STUDIES ON THE INSTRUMENTAL FACTORS DETERMINING ERRORS IN THE MEASUREMENTS OF RAMAN BAND PROFILES .1. OPTICAL FACTORS - A COMPARISON OF 3 DIFFERENT METHODS

Citation
L. Mariani et al., STUDIES ON THE INSTRUMENTAL FACTORS DETERMINING ERRORS IN THE MEASUREMENTS OF RAMAN BAND PROFILES .1. OPTICAL FACTORS - A COMPARISON OF 3 DIFFERENT METHODS, Applied spectroscopy, 47(8), 1993, pp. 1227-1233
Citations number
20
Categorie Soggetti
Instument & Instrumentation",Spectroscopy
Journal title
ISSN journal
00037028
Volume
47
Issue
8
Year of publication
1993
Pages
1227 - 1233
Database
ISI
SICI code
0003-7028(1993)47:8<1227:SOTIFD>2.0.ZU;2-E
Abstract
A treatnient is proposed which takes into account the following error sources in Raman band shape measurements: incident beam polarization, polarization analyzer misorientation, monochromator response, and obse rvation over a finite solid angle. Three methods generally adopted for this type of measurements are compared in terms of the amount of spur ious anisotropic component which enters the isotropic profile. The per tinence zones of each method are shown. The percentage of error is exp ressed by well-defined quantities entering analytical equations, this allows the experimentalist to choose the more suitable method on the b asis of the experimental setup.