TIME-OF-FLIGHT MASS REFLECTRON WITH A LARGE-AREA OF ION COLLECTION

Citation
Dr. Bandura et Aa. Makarov, TIME-OF-FLIGHT MASS REFLECTRON WITH A LARGE-AREA OF ION COLLECTION, International journal of mass spectrometry and ion processes, 127, 1993, pp. 45-55
Citations number
24
Categorie Soggetti
Spectroscopy,"Physics, Atomic, Molecular & Chemical
ISSN journal
01681176
Volume
127
Year of publication
1993
Pages
45 - 55
Database
ISI
SICI code
0168-1176(1993)127:<45:TMRWAL>2.0.ZU;2-F
Abstract
This paper is devoted to the construction of a time-of-flight mass ref lectron with the following features: (1) a time resolution of several hundreds for various ion sources (providing elemental analysis with su rface ionization, Cf-252 or dust impact ion sources); (2) a huge area of the ion formation region (200 cm2) with the possibility of extendin g further without decreasing analytical characteristics or lengthening the reflectron); (3) an extremely simplified compact design with very high transparency of the ion mirror grid (> 98%) and decreased demand s for accuracy of manufacturing and voltage supply. An experimentally observed satisfactory agreement between the theoretical (both analytic al and numerical) model of the reflectron and its practical realizatio n, which is made possible by taking into account large-scale field per turbations, caused by grids on the ion mirror and the secondary electr on multiplier, as well as by near-emitter chromatic aberration.