Mh. Davis et al., ASSESSMENT OF LEFT-VENTRICULAR EJECTION FRACTION FROM TECHNETIUM-99M-METHOXY ISOBUTYL ISONITRILE MULTIPLE-GATED RADIONUCLIDE ANGIOCARDIOGRAPHY, IEEE transactions on medical imaging, 12(2), 1993, pp. 189-199
Citations number
32
Categorie Soggetti
Engineering, Biomedical","Radiology,Nuclear Medicine & Medical Imaging
Abnormal left ventricular function is a diagnostic indication of cardi
ac disease. Left ventricular function is commonly quantified by ejecti
on fraction measurements. A novel approach for the determination of le
ft ventricular ejection fraction from technetium-99m-methoxy isobutyl
isonitrile multiple-gated radionuclide angiocardiography is presented.
Data from 23 patients, symptomatic of cardiac disease, indicate that
ejection fractions determined using the radionuclide technique correla
te well with contrast X-ray single-plane cineangiography (r = 0.83, p
< 10(-6)). Data from 14 of the patients indicate favorable correlation
with technetium-99m-pertechnetate gated blood pool radionuclide angio
cardiography (r = 0.87, p < 10(-4)).