W. Seiler et C. Hiemke, DETERMINATION OF HOMOVANILLIC-ACID IN HUMAN PLASMA USING HPLC WITH ELECTROCHEMICAL DETECTION AND AUTOMATED SOLID-PHASE EXTRACTION, Journal of liquid chromatography, 16(13), 1993, pp. 2813-2823
An isocratic HPLC method with electrochemical detection for the quanti
fication of homovanillic acid (HVA) is described. The method included
automated solid phase extraction on C18-reversed phase material, follo
wed by separation on a 3-mum Nucleosil 100 C18 column (250 mm x 4.6 mm
I.D.) with a 100 mM citric acid solution (pH 6.6) containing 4% aceto
nitrile (v/v) as eluent at a flow rate of 0.6 ml/min. Isovanillinic ac
id served as internal standard. Extractability of both analytes was ca
. 80 %. After extraction of 1 ml of plasma, coefficients of variation
of replicate analyses were below 10 % in the naturally occuring concen
tration range.