AN ELECTROTHERMAL MODEL FOR LARGE MASS BOLOMETRIC DETECTORS

Citation
A. Alessandrello et al., AN ELECTROTHERMAL MODEL FOR LARGE MASS BOLOMETRIC DETECTORS, IEEE transactions on nuclear science, 40(4), 1993, pp. 649-656
Citations number
10
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
40
Issue
4
Year of publication
1993
Part
1
Pages
649 - 656
Database
ISI
SICI code
0018-9499(1993)40:4<649:AEMFLM>2.0.ZU;2-B
Abstract
A method for modelling large mass bolometric detectors was developed. It is based on the analysis of the different pulse shapes obtained whe n a particle deposits its energy in the absorber or in the temperature sensor. SPICE was used for the simulation. Small and large signal ana lysis is carried on. The model was used for the accurate parameter ext raction of two TeO2 detectors of 70g and 330g respectively. Results su ggest a possible field-effect dependence of the electron-phonon therma l conductance and electron heat capacity in the NTD thermistors used.