Ccp. Eagle et Df. Capes, USE OF A NEW SYRINGE PUMP (SPRINGFUSOR(R)) FOR MUSCLE-RELAXANT INFUSION, Anaesthesia and intensive care, 21(4), 1993, pp. 444-446
Citations number
6
Categorie Soggetti
Anesthesiology,"Emergency Medicine & Critical Care
A new spring-driven syringe pump (Springfusor 10(R), Go Medical Indust
ries, Subiaco, W.A.) was evaluated for infusion of muscle relaxants in
50 surgical cases. After an initial bolus dose, atracurium was given
at a rate of 25 to 37.5 mg/hour and vecuronium at a rate of 4 to 6 mg/
hour. The mean infusion rate was found to be 4% above the nominal rate
of 5 ml/hour and the accuracy in all cases was within +/- 20%. The Sp
ringfusor was found to be rugged, cheap and easy to use. It lacks dose
flexibility but is suitable for continuous background infusion of mus
cle relaxants during surgery.