USE OF A NEW SYRINGE PUMP (SPRINGFUSOR(R)) FOR MUSCLE-RELAXANT INFUSION

Citation
Ccp. Eagle et Df. Capes, USE OF A NEW SYRINGE PUMP (SPRINGFUSOR(R)) FOR MUSCLE-RELAXANT INFUSION, Anaesthesia and intensive care, 21(4), 1993, pp. 444-446
Citations number
6
Categorie Soggetti
Anesthesiology,"Emergency Medicine & Critical Care
ISSN journal
0310057X
Volume
21
Issue
4
Year of publication
1993
Pages
444 - 446
Database
ISI
SICI code
0310-057X(1993)21:4<444:UOANSP>2.0.ZU;2-5
Abstract
A new spring-driven syringe pump (Springfusor 10(R), Go Medical Indust ries, Subiaco, W.A.) was evaluated for infusion of muscle relaxants in 50 surgical cases. After an initial bolus dose, atracurium was given at a rate of 25 to 37.5 mg/hour and vecuronium at a rate of 4 to 6 mg/ hour. The mean infusion rate was found to be 4% above the nominal rate of 5 ml/hour and the accuracy in all cases was within +/- 20%. The Sp ringfusor was found to be rugged, cheap and easy to use. It lacks dose flexibility but is suitable for continuous background infusion of mus cle relaxants during surgery.