Cp. Montoya et al., AN ELECTRON-POSITRON PAIR SPECTROMETER FOR HIGH-ENERGY DECAYS OF NUCLEI, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 334(2-3), 1993, pp. 437-446
A highly segmented phoswich array of plastic scintillators has been co
nstructed for measurements of (e+, e-) pairs emitted in high energy el
ectromagnetic transitions in nuclei. Electron (positron) energies of 2
-30 MeV can be measured by each individual element, with a total trans
ition energy resolution of deltaE/E = 13% for a 20 MeV transition. The
array covers 29% of 4pi and its efficiency is 1.6% for a 6 MeV E0 int
ernal pair decay, and 1.1% for an 18 MeV El transition. These efficien
cy values include cuts which reduce background and optimize energy res
olution. Experimental results are compared with GEANT simulations demo
nstrating the detector's ability to determine transition multipolariti
es.