Hp. Cresswell, EVALUATION OF THE PORTABLE PRESSURE TRANSDUCER TECHNIQUE FOR MEASURING FIELD TENSIOMETERS, Australian Journal of Soil Research, 31(4), 1993, pp. 397-406
The use of a portable pressure transducer system which couples to tens
iometers via a syringe needle and septum stopper was assessed. Changes
that occurred to pressure within the tensiometer on insertion of the
needle were significant and up to 120 min was sometimes required befor
e complete re-equilibration to the original tensiometric pressure was
achieved. Tensiometer readings with less than 5% error were attained b
y allowing 3-4 min re-equilibration time at soil matric potentials bet
ween -1 and -25 kPa, but up to 35 min was required at -40 kPa. As well
as increasing with decreasing initial pressure in the tensiometer, re
-equilibration time was dependent on soil properties. The re-equilibra
tion times reported are specific to the volume of air left in the tens
iometer below the septum stopper (1.2 cm3). A larger air bubble will d
ecrease re-equilibration time but could cause undesirable thermal erro
rs during field use. Transducer readings should be made after allowing
time for complete re-equilibration of the tensiometer following inser
tion of the needle. However, for practical field measurement of low ma
tric potentials, it will be necessary to read the portable transducer
before full re-equilibration and correct the reading based on previous
ly determined re-equilibration patterns. These re-equilibration patter
ns should be specific to each tensiometer and should cover a range of
matric potentials. It is also important that the air bubble below the
septum stopper be maintained constant for each reading, and that the s
eptum stoppers be replaced regularly. With careful use, these systems
are convenient for monitoring large numbers of tensiometers spaced apa
rt in the field.