Dd. Deshpande et al., DOSIMETRY OF INTRACAVITARY APPLICATIONS IN CARCINOMA OF THE CERVIX - RECTAL DOSE ANALYSIS, Radiotherapy and oncology, 42(2), 1997, pp. 163-166
Citations number
7
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
The International Commission on Radiation Units and Measurements (ICRU
), in its Report-38, has given certain recommendations regarding the s
pecifications of bladder and rectal reference points in the intracavit
ary treatment of carcinoma of the cervix. Conforming to this report, 1
82 intracavitary applications performed in stage I and II cervix cases
were analyzed. In 113 applications, the maximum rectal dose was found
to be in a point on the anteroposterior line drawn through the centre
of colpostat sources. However, for our type of applications, the poin
t on the anteroposterior line drawn through the lower end of the uteri
ne tube seldom gets maximum dose. In addition, it was observed that th
ere are other dose points than the ICRU reference point receiving dose
s close to maximum dose. It was concluded that doses to multiple recta
l reference points should be recorded, in addition to the ICRU-defined
rectal reference point. (C) 1997 Elsevier Science Ireland Ltd.