MUTUAL COMPARISONS OF FULLY AUTOMATED JOSEPHSON-JUNCTION ARRAY VOLTAGE-STANDARD SYSTEMS

Citation
A. Iwasa et al., MUTUAL COMPARISONS OF FULLY AUTOMATED JOSEPHSON-JUNCTION ARRAY VOLTAGE-STANDARD SYSTEMS, IEEE transactions on instrumentation and measurement, 42(2), 1993, pp. 368-371
Citations number
3
Categorie Soggetti
Engineering, Eletrical & Electronic","Instument & Instrumentation
ISSN journal
00189456
Volume
42
Issue
2
Year of publication
1993
Pages
368 - 371
Database
ISI
SICI code
0018-9456(1993)42:2<368:MCOFAJ>2.0.ZU;2-#
Abstract
In order to test the performances of three sets of fully automated Jos ephson-junction array voltage-standard systems, mutual comparisons amo ng those systems have been carried out. The results obtained show that each system can be reasonably operated with the assigned uncertainty of about one part in 10(8) in routine daily calibration of Zener-diode voltage standards.