ACCURATE DETERMINATION OF THE DENSITY OF A CRYSTAL SILICON SPHERE

Citation
K. Fujii et al., ACCURATE DETERMINATION OF THE DENSITY OF A CRYSTAL SILICON SPHERE, IEEE transactions on instrumentation and measurement, 42(2), 1993, pp. 395-400
Citations number
30
Categorie Soggetti
Engineering, Eletrical & Electronic","Instument & Instrumentation
ISSN journal
00189456
Volume
42
Issue
2
Year of publication
1993
Pages
395 - 400
Database
ISI
SICI code
0018-9456(1993)42:2<395:ADOTDO>2.0.ZU;2-G
Abstract
For an independent determination of the Avogadro constant, the density of a 1-kg crystal silicon sphere was determined by direct measurement s of its mass and volume. A scanning type interferometer was used to m easure the diameters, and the volume was calculated from the mean of u niformly distributed diameters. The sphere was weighed using a balance for the Prototype Kilogram at the NRLM, and the mass was determined a ccurately by direct measurements of the buoyancy force acting on the s phere. The total uncertainty of the density is estimated to be 0.34 pp m. Effects of a thin oxide layer and impurities on the density of a pu re crystal are evaluated.