Jr. Kinard et al., INTERCOMPARISON OF NIST, NPL, PTB, AND VSL THERMAL VOLTAGE CONVERTERSFROM 100 KHZ TO 1-MHZ, IEEE transactions on instrumentation and measurement, 42(2), 1993, pp. 615-617
Coaxial, thermal voltage converters were hand-carried among NIST, NPL,
PTB, and VSL for intercomparison of ac-dc difference from 100 kHz to
1 MHz. This paper briefly describes the highly varied methods and unde
rlying principles on which ac-dc difference determinations are based i
n each laboratory, the transport standards used, and the results of th
e intercomparisons. The ac-dc differences reported by the participatin
g laboratories are in very good agreement. It has been concluded that
representative values of ac-dc difference in the 100 kHz to 1 MHz freq
uency range have been established with 2sigma uncertainties of 4-8 ppm
.