DETECTOR RESPONSE AND EXPOSURE CONTROL OF THE RADIO VISIOGRAPHY SYSTEM (RVG 32000 ZHR)

Citation
Sk. Chen et al., DETECTOR RESPONSE AND EXPOSURE CONTROL OF THE RADIO VISIOGRAPHY SYSTEM (RVG 32000 ZHR), Oral surgery, oral medicine, oral pathology, 76(1), 1993, pp. 104-111
Citations number
12
Categorie Soggetti
Pathology,Surgery,"Dentistry,Oral Surgery & Medicine
ISSN journal
00304220
Volume
76
Issue
1
Year of publication
1993
Pages
104 - 111
Database
ISI
SICI code
0030-4220(1993)76:1<104:DRAECO>2.0.ZU;2-T
Abstract
The RadioVisioGraphy system (RVG 32000 ZHR) was studied with respect t o some physical properties including the precision and linearity of th e exposure control and the detector response to x-rays. The x-ray unit was found to yield repetitive exposures that are precisely controlled by the CCX timer. A ''preheat'' and an ''after exposure'' period were noticed, and their resulting x-ray exposures were measured. Two RVG d etectors were compared for their responses with x-rays. The difference s between the detectors were insignificant. However, inhomogeneous det ector responses were found in both of the detectors; with the same exp osure, the central parts of the detectors yielded lower pixel values t han the peripheral parts. A calibration equation was developed to corr ect for the inhomogeneity.