Xc. Chen et al., THE MEASUREMENT OF BACKSCATTER COEFFICIENT FROM A BROAD-BAND PULSE-ECHO SYSTEM - A NEW FORMULATION, IEEE transactions on ultrasonics, ferroelectrics, and frequency control, 44(2), 1997, pp. 515-525
A new formulation for obtaining the absolute backscatter coefficient f
rom pulse-echo measurements is presented. Using this formulation, perf
orming the diffraction correction and system calibration is straightfo
rward. The diffraction correction function for the measurement of back
scatter coefficient and the acoustic coupling function for a pulse-ech
o system are defined. Details of these functions for two very useful c
ases are presented: a flat disk transducer and a spherically focused t
ransducer. Approximations of these functions are also provided. For a
flat disk transducer, the final formulation appears as a modification
to the established Sigelmann-Reid formulation. For a focused transduce
r, the final correction is a weak function of frequency when the scatt
ering volume is near the focal area, rather than the frequency squared
dependence proposed by earlier investigators.