DOSE CONTROL IN ELECTRON-BEAM PROCESSING - COMPARISON OF RESULTS FROMA GRAPHITE CHARGE COLLECTOR, ROUTINE DOSIMETERS AND THE ISS ALANINE-BASED DOSIMETER
Pg. Fuochi et al., DOSE CONTROL IN ELECTRON-BEAM PROCESSING - COMPARISON OF RESULTS FROMA GRAPHITE CHARGE COLLECTOR, ROUTINE DOSIMETERS AND THE ISS ALANINE-BASED DOSIMETER, Radiation physics and chemistry, 42(4-6), 1993, pp. 809-812
Citations number
7
Categorie Soggetti
Nuclear Sciences & Tecnology","Chemistry Physical","Physics, Atomic, Molecular & Chemical
A 12 MeV linear accelerator is currently used for electron beam proces
sing of power semiconductor devices for lifetime control and, on an ex
perimental basis, for food irradiation, sludge treatment etc. In order
to control tbe irradiation process a simple, quick and reliable metho
d for a direct evaluation of dose and fluence in a broad electron beam
has been developed. This paper presents the results obtained using a
'charge collector' which measures the charge absorbed in a graphite ta
rget exposed in air. Calibration of the system with super-Fricke dosim
eter and comparison of absorbed dose results obtained with plastic dos
imeters and alanine pellets are discussed.