Fx. Chen et al., STUDIES AND IMPROVEMENT OF RADIATION-INDUCED COLOR-CENTER DOSIMETER IN LIF, Radiation physics and chemistry, 42(4-6), 1993, pp. 833-835
Citations number
5
Categorie Soggetti
Nuclear Sciences & Tecnology","Chemistry Physical","Physics, Atomic, Molecular & Chemical
We have studied the properties of the dosimeter which was made of home
manufactured LiF crystal. The crystal was cut into chips and each of w
hich was 6 mm X 6 mm X 1 mm in size. The value of absorbed dose in LiF
was calibrated by alanine/ESR dosimetry. The alanine samples were sup
plied and measured by National Institute of Metrology. The irradiation
s were carried out with a cobalt-60 source of 7.4 X 10(15) Bq of Tianj
in TechnicaL Physics Institute. The dose-effects of absorbance peaks f
or F,N,R and N band, in specific dose ranges, can all be expressed as
A=bD(n), where A is the increase in average optical absorbance; b and
n are constant relating to the type of color-centers and the impurity
of the crystal. We also study to anneal the crystal after irradiation,
stable time is cut down from 24hr. to 4hr.