USES OF POLYMER-ALANINE FILM ESR DOSIMETERS IN DOSIMETRY OF IONIZING-RADIATION

Citation
Lq. Xie et al., USES OF POLYMER-ALANINE FILM ESR DOSIMETERS IN DOSIMETRY OF IONIZING-RADIATION, Radiation physics and chemistry, 42(4-6), 1993, pp. 837-840
Citations number
3
Categorie Soggetti
Nuclear Sciences & Tecnology","Chemistry Physical","Physics, Atomic, Molecular & Chemical
ISSN journal
0969806X
Volume
42
Issue
4-6
Year of publication
1993
Pages
837 - 840
Database
ISI
SICI code
0969-806X(1993)42:4-6<837:UOPFED>2.0.ZU;2-L
Abstract
Alanine ESR dosimetry is a reliable method, used in a various fields o f ionizing radiation. The polymer-alanine film/ESR dosimeters of 0.3 - 0.4 mm thickness were prepared and their dosimetric properties were s tudied for Co-60 gamma photons and 3 - 5 MeV electrons in the dose ran ge from 20 Gy to 100 kGy. The results show that under normal condition s the alanine calibration curves are linear in the dose range from 100 Gy to 10kGy. The dose profiles at the electron radiation field were m easured with the film alanine dosimeters. The polymer-alanine film dos imeters were used for ion implantation of 400 keV ion implantor. Their dose response and energy dependence were investigated initially.