PERFUSION IMAGING BY UN-INVERTED FLOW-SENSITIVE ALTERNATING INVERSION-RECOVERY (UNFAIR)

Citation
Ja. Helpern et al., PERFUSION IMAGING BY UN-INVERTED FLOW-SENSITIVE ALTERNATING INVERSION-RECOVERY (UNFAIR), Magnetic resonance imaging, 15(2), 1997, pp. 135-139
Citations number
9
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
0730725X
Volume
15
Issue
2
Year of publication
1997
Pages
135 - 139
Database
ISI
SICI code
0730-725X(1997)15:2<135:PIBUFA>2.0.ZU;2-W
Abstract
A new pulse sequence for estimating cerebral blood flow called UNFAIR, which uses a combination of sequential hyperbolic secant preparatory pulses, is introduced. This sequence is based on the same generalized conditions as previously introduced inversion recovery techniques exce pt that the spins in the image slice of interest always have +z magnet ization and the in-flowing spins are alternately inverted and uninvert ed. CBF-weighted images of rat brain under conditions of normocpnia an d hypercapnia are presented and demonstrate the expected CBF response. A model describing the signal response to this pulse sequence is also presented and compared with in-vivo data acquired from gray and white matter. (C) 1997 Elsevier Science Inc.