CALIBRATION OF THE NPL SECONDARY STANDARD RADIONUCLIDE CALIBRATOR FORIR-192 BRACHYTHERAPY SOURCES

Citation
Jp. Sephton et al., CALIBRATION OF THE NPL SECONDARY STANDARD RADIONUCLIDE CALIBRATOR FORIR-192 BRACHYTHERAPY SOURCES, Physics in medicine and biology, 38(8), 1993, pp. 1157-1164
Citations number
10
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
38
Issue
8
Year of publication
1993
Pages
1157 - 1164
Database
ISI
SICI code
0031-9155(1993)38:8<1157:COTNSS>2.0.ZU;2-F
Abstract
Safe and effective treatment with brachytherapy sources requires an ac curate knowledge of the local tissue absorbed dose rate derived from t he source reference air kerma rate. It is desirable that these air ker ma rate measurements be traceable to national standards. The NPL has e mbarked on a programme that will enable the user to assay brachytherap y sources in a convenient manner prior to treatment. Calibration figur es have been derived for the NPL secondary standard radionuclide calib rator for Ir-192 brachytherapy sources manufactured by Amersham Intern ational plc. The calibration figures enable the user to accurately est imate the reference air kerma rate and activity of such sources by mea suring the ionization chamber response. Calibration figures for other brachytherapy sources are also being derived.