HIGH-PRESSURE, HIGH-TEMPERATURE 2-CHAMBER CELL WITH CHANGEABLE PATH LENGTHS FOR ACCURATE MEASUREMENTS OF ABSORPTION-COEFFICIENT

Citation
Ye. Gorbaty et Gv. Bondarenko, HIGH-PRESSURE, HIGH-TEMPERATURE 2-CHAMBER CELL WITH CHANGEABLE PATH LENGTHS FOR ACCURATE MEASUREMENTS OF ABSORPTION-COEFFICIENT, Review of scientific instruments, 64(8), 1993, pp. 2346-2349
Citations number
5
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
64
Issue
8
Year of publication
1993
Pages
2346 - 2349
Database
ISI
SICI code
0034-6748(1993)64:8<2346:HH2CWC>2.0.ZU;2-6
Abstract
A cell with two sample chambers arranged in sequence is designed to pr ovide accurate absorption coefficients for the components of binary fl uid systems at high pressures and temperatures. Once primed, the cell allows one to obtain results over the whole range of temperature and p ressure required without disassembly and refilling. The method elimina tes or at least diminishes most of the possible errors automatically.