A. Bouchalkha et al., DESIGN CHANGES IN THE SAMPLE MOUNT AND TEMPERATURE SENSORS OF THE JANIS SUPERVARITEMP DEWAR - LIQUID-NITROGEN OPERATION, Review of scientific instruments, 64(8), 1993, pp. 2389-2390
A new sample holder for the Janis Supervaritemp Dewar has been designe
d which allows very accurate temperature readings at the sample and av
oids sample breakage by reducing thermally induced stress. In addition
, the successful operation of this helium dewar using liquid nitrogen
has been achieved allowing temperatures above 80 K, stable to within /-0.04 K, to be obtained.