J. Vanderweide et al., EVALUATION OF THE CLONED ENZYME DONOR IMMUNOASSAY FOR MEASUREMENT OF PHENYTOIN AND PHENOBARBITAL IN SERUM, Therapeutic drug monitoring, 15(4), 1993, pp. 344-348
We report the evaluation of the cloned enzyme donor immunoassay (CEDIA
) for the estimation of phenytoin and phenobarbital in serum. The assa
ys were performed with a Hitachi 911 analyzer. Intra-assay coefficient
s of variation were from 1.5 to 4.4% for phenytoin and 1.6 to 5.5% for
phenobarbital. Interassay coefficients of variation ranged from 1.8 t
o 5.3% for phenytoin and 2.9 to 4.8% for phenobarbital. Linearity was
satisfactory, with a recovery of 103% at 11 mg/L and 110% at 22 mg/L p
henytoin and 93% at 14 mg/L and 101 % at 40 mg/L phenobarbital. The de
tection limit was 1.2 mg/L for phenytoin and 0.6 mg/L for phenobarbita
l. Results of this assay correlated well with those of a conventional
high-performance liquid chromatography (HPLC) method: r = 0.99 for phe
nytoin (n = 47) and r = 0.98 for phenobarbital (n = 48). The CEDIA was
easy to handle and especially suitable for short turn-around time app
lication.