EVALUATION OF THE CLONED ENZYME DONOR IMMUNOASSAY FOR MEASUREMENT OF PHENYTOIN AND PHENOBARBITAL IN SERUM

Citation
J. Vanderweide et al., EVALUATION OF THE CLONED ENZYME DONOR IMMUNOASSAY FOR MEASUREMENT OF PHENYTOIN AND PHENOBARBITAL IN SERUM, Therapeutic drug monitoring, 15(4), 1993, pp. 344-348
Citations number
10
Categorie Soggetti
Pharmacology & Pharmacy","Public, Environmental & Occupation Heath",Toxicology,Biology
Journal title
ISSN journal
01634356
Volume
15
Issue
4
Year of publication
1993
Pages
344 - 348
Database
ISI
SICI code
0163-4356(1993)15:4<344:EOTCED>2.0.ZU;2-Q
Abstract
We report the evaluation of the cloned enzyme donor immunoassay (CEDIA ) for the estimation of phenytoin and phenobarbital in serum. The assa ys were performed with a Hitachi 911 analyzer. Intra-assay coefficient s of variation were from 1.5 to 4.4% for phenytoin and 1.6 to 5.5% for phenobarbital. Interassay coefficients of variation ranged from 1.8 t o 5.3% for phenytoin and 2.9 to 4.8% for phenobarbital. Linearity was satisfactory, with a recovery of 103% at 11 mg/L and 110% at 22 mg/L p henytoin and 93% at 14 mg/L and 101 % at 40 mg/L phenobarbital. The de tection limit was 1.2 mg/L for phenytoin and 0.6 mg/L for phenobarbita l. Results of this assay correlated well with those of a conventional high-performance liquid chromatography (HPLC) method: r = 0.99 for phe nytoin (n = 47) and r = 0.98 for phenobarbital (n = 48). The CEDIA was easy to handle and especially suitable for short turn-around time app lication.