Jf. Vanstaden et M. Matoetoe, DETERMINATION OF COPPER BY ANODIC-STRIPPING VOLTAMMETRY ON A GLASSY-CARBON ELECTRODE USING A CONTINUOUS-FLOW SYSTEM, Fresenius' journal of analytical chemistry, 357(6), 1997, pp. 624-628
A successful flow-through system was developed for trace analysis of c
opper using DPASV with a glassy-carbon electrode. Periodical chemical
regeneration of the electrode with a 1 mol/L NaOH solution increased s
ensitivity and precision. The method was shown to be applicable with a
detection limit of 0.56 mu g/L, with a determination time of less tha
n 7 min per measurement (without deaeration time). The draw-back of th
e system is the 10 min deaeration time. The system gave an accuracy of
0.090 +/- 0.005% for a certified reference material of low alloy stee
l containing 0.090 +/- 0.004% Cu. Applicability to various fresh water
samples with a Cu content between 1.57 and 13.11 mu g/L with an RSD <
2.36% is illustrated.