S. Avramov et Nm. Oldham, AUTOMATIC CALIBRATION OF INDUCTIVE VOLTAGE DIVIDERS FOR THE NASA ZENOEXPERIMENT, Review of scientific instruments, 64(9), 1993, pp. 2676-2678
Two inductive voltage dividers (IVDs) used for temperature measurement
s in NASA's Zeno experiment were tested. In order to obtain the requir
ed resolution of 10 parts-per-billion, a 30-bit binary inductive volta
ge divider developed at the National Institute of Standards and Techno
logy was used to measure the differential linearity of the Zeno IVDs.
Automatic measurements were performed on the dividers in the Zeno engi
neering model at frequencies of 266 and 351 Hz over a ratio range of 0
.55-0.56. The measured differential linearity limits the temperature r
esolution to 5 muK.