Bm. Chien et al., ENHANCEMENT OF RESOLUTION IN MATRIX-ASSISTED LASER-DESORPTION USING AN ION-TRAP STORAGE REFLECTRON TIME-OF-FLIGHT MASS-SPECTROMETER, Rapid communications in mass spectrometry, 7(9), 1993, pp. 837-843
A quadrupole ion-trap storage/reflectron time-of-flight mass spectrome
ter (IT/RETOFMS) has been used to enhance resolution in the ultraviole
t matrix-assisted laser desorption/ionization (MALDI) of peptides. The
MALDI of peptides is performed directly inside an ion-trap storage de
vice using 266 nm laser radiation from a 2,5-dihydroxybenzoic acid mat
rix. Helium buffer gas is used to collisionally cool the hot desorbed
ions and to rapidly relax the ions to the center of the ion trap. The
result is that upon ejection into the TOF device, a resolution of near
ly 6200 is obtained for Substance P using a 90 ms storage time. A reso
lution of 3000-6000 is consistently obtained for peptides up to 2100 u
used in this study. The individual ion peaks from the isotopic distri
bution are easily resolved under these conditions. In addition, the io
n trap is found to be useful as an ion-storage device for enhancing th
e signal where multiple laser desorption pulses can be stored per cycl
e in the trap before ejection into the RETOF. Also, using the relative
ly high resolving power of the IT/RETOF instrument, excellent mass acc
uracy can be obtained using the matrix background and bradykinin as in
ternal standards.