P. Dobberstein et E. Schroeder, ACCURATE MASS DETERMINATION OF A HIGH-MOLECULAR-WEIGHT PROTEIN USING ELECTROSPRAY-IONIZATION WITH A MAGNETIC-SECTOR INSTRUMENT, Rapid communications in mass spectrometry, 7(9), 1993, pp. 861-864
Accurate mass determination on the 13+ ion of myoglobin, MW 16 951, ha
s been performed with a magnetic sector mass spectrometer equipped wit
h a newly designed electrospray ion source. At a mass resolution of 20
000 a mass deviation of 1 ppm from the expected value has been achiev
ed.