MTF EVALUATION OF A PHOSPHOR-COATED CCD FOR X-RAY-IMAGING

Citation
M. Gambaccini et al., MTF EVALUATION OF A PHOSPHOR-COATED CCD FOR X-RAY-IMAGING, Physics in medicine and biology, 41(12), 1996, pp. 2799-2806
Citations number
8
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
41
Issue
12
Year of publication
1996
Pages
2799 - 2806
Database
ISI
SICI code
0031-9155(1996)41:12<2799:MEOAPC>2.0.ZU;2-2
Abstract
A novel digital x-ray detector has been assembled. The imaging system is based on a phosphor-coated charge-coupled device (CCD) obtained by direct deposition of a gadolinium oxysulphide scintillator onto the de tector surface. The modulation transfer function has been measured alo ng the two directions of the digital coordinates with the narrow slit technique. A resolution limit of about 20 line pairs per mm has been o btained for both directions. The high spatial resolution currently dem anded in mammography can be achieved with this imaging system.