AN EVALUATION OF THE SENS-A-RAY DIGITAL DENTAL IMAGING-SYSTEM

Citation
D. Mcdonnell et C. Price, AN EVALUATION OF THE SENS-A-RAY DIGITAL DENTAL IMAGING-SYSTEM, Dento-maxillo-facial radiology, 22(3), 1993, pp. 121-126
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
0250832X
Volume
22
Issue
3
Year of publication
1993
Pages
121 - 126
Database
ISI
SICI code
0250-832X(1993)22:3<121:AEOTSD>2.0.ZU;2-Q
Abstract
The Sens-A-Ray digital imaging system was compared with group D- and E -speed intraoral radiographic film with respect to resolution, speed a nd image quality. Using a line-pair plate, resolution was determined t o be 10 line-pairs per millimetre. With a constant potential X-ray sou rce, the speed of the system was over three times that of E-speed emul sion and the response of pixel value to exposure was linear. Absorptio n curves with aluminium showed a similar response to D-speed film. An aluminium foil test object was used to compare the Sens-A-Ray system w ith both film types. Thirty observers were scored for successful ident ification of a pattern of holes in the test object. There was little d ifference between D- and E-speed film groups, and successful identific ation was significantly reduced with the Sens-A-Ray images when compar ed with D- and E-speed films.