FAST DETERMINATION OF WHOLE-BLOOD GLUCOSE WITH A CALORIMETRIC MICRO-BIOSENSOR

Citation
B. Xie et al., FAST DETERMINATION OF WHOLE-BLOOD GLUCOSE WITH A CALORIMETRIC MICRO-BIOSENSOR, Sensors and actuators. B, Chemical, 15(1-3), 1993, pp. 141-144
Citations number
8
Categorie Soggetti
Engineering, Eletrical & Electronic","Instument & Instrumentation
ISSN journal
09254005
Volume
15
Issue
1-3
Year of publication
1993
Pages
141 - 144
Database
ISI
SICI code
0925-4005(1993)15:1-3<141:FDOWGW>2.0.ZU;2-J
Abstract
Fast determination of whole blood glucose without any pretreatment was achieved using a calorimetric FIA micro-biosensor. Glucose oxidase wa s covalently immobilized, together with catalase, onto controlled pore glass beads and packed into a micro-column. By reducing the sample vo lume down to 1 mul, a measurement period of about 40 s and an enlarged linear range of 0.5-20 mM glucose were obtained. The cor-relation coe fficient was 0.98 when the micro-biosensor was compared to the Reflolu x-S blood glucose analyser. The relative standard deviation for 100 bl ood samples was 3.7%.