MEASUREMENT OF OXYGEN PARTIAL-PRESSURE USING FULLY-SEALED ZIRCONIA PUMP GAUGE DEVICES OPERATED IN THE TRACKING MODE .2.

Citation
M. Benammar et Wc. Maskell, MEASUREMENT OF OXYGEN PARTIAL-PRESSURE USING FULLY-SEALED ZIRCONIA PUMP GAUGE DEVICES OPERATED IN THE TRACKING MODE .2., Sensors and actuators. B, Chemical, 15(1-3), 1993, pp. 162-165
Citations number
9
Categorie Soggetti
Engineering, Eletrical & Electronic","Instument & Instrumentation
ISSN journal
09254005
Volume
15
Issue
1-3
Year of publication
1993
Pages
162 - 165
Database
ISI
SICI code
0925-4005(1993)15:1-3<162:MOOPUF>2.0.ZU;2-U
Abstract
The instrument incorporated a miniature oxygen sensor operated in an a .c. mode. Signal processing of the periodic gauge e.m.f. enabled deter mination of the oxygen partial pressures both within the enclosed volu me of the sensor (P0) and in the surrounding sample gas (P1). The devi ce was operated in a closed loop in order to control P0. The mean gaug e e.m.f. was maintained equal to zero resulting in a well-defined rela tion between P0 and P1. P1 was then computed from a measurement of the amplitude of the gauge e.m.f., determined using simple circuity. Furt hermore, theory indicated that linearity of the output should be maint ained at high values of the amplitude of the gauge e.m.f. resulting in high signal-to-noise ratio: this was confirmed for peak-to-peak value s up to 50 mV.