Of. Naismith et Dj. Thomas, A PERFORMANCE-TEST OF UK PERSONAL DOSIMETRY SERVICES FOR FAST-NEUTRONS, Radiation protection dosimetry, 48(4), 1993, pp. 333-340
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
In order to provide quantitative information on the quality of fast ne
utron dosimetry services in the UK, the NPL has carried out a performa
nce test of these services. A summary is given of the dosemeters teste
d, and the irradiation methods are described. Neutrons from both an Am
-214-Be and a Cf-252 source were used to deliver dose equivalents in t
he range 0.5 mSv to 170 mSv. The data have been analysed in terms of t
he bias and standard deviation in the ratios of reported to reference
dose equivalent. There was a tendency for dosemeters to over-respond a
t the lower dose equivalent values and under-respond at higher values.
The spread in individual readings was high for the lowest dose equiva
lents, but decreased as the dose equivalent became larger.