F. Mertzlufft et R. Zander, PERIOPERATIVE RESPIRATORY MONITORING OF OXYGEN-TRANSPORT, Infusionstherapie und Transfusionsmedizin, 20(4), 1993, pp. 180-184
Oxymetry nowadays is understood as the in vitro measurement of O2 satu
ration (sO2, %) and hemoglobin (Hb) derivatives (%) using 4-7 waveleng
ths (CO- and Hem-oxymeters). Pulse oxymeters, using only 2 wavelengths
, are designed for the continuous noninvasive measurement of the arter
ial partial O2 saturation (psO2, %) in vivo. Light-emitting diodes all
ow light to pass through the peripheral site of measurement with red a
nd infrared light to enable a distinction between oxygenated and deoxy
genated hemoglobin during a recorded pulse wave. In case of physiologi
cal concentrations of Hb derivatives the determination of psO2 is perf
ormed with clinically relevant accuracy of +/- 2-3%. However, at carbo
xyhemoglobin (COHb) concentrations above normal, under normoxia as wel
l as under hypoxia the accuracy of measurement varies considerably amo
ng the instruments from different manufacturers. In the case of elevat
ed methemoglobin (MetHb) concentrations, the situation is completely d
ifferent. With increasing cMetHb, the psO2 is still the value required
, but success depends on the concentration of MetHb: under normoxia ps
O2 is increasingly underestimated, whereas under hypoxia increasing ov
erestimation must be anticipated. Provided there is a constant Hb conc
entration, knowledge of the initial sO2, and absence of the derivative
s COHb and MetHb as well as of severe perfusion disorders, pulse oxyme
try is suitable for perioperative respiratory monitoring of oxygen tra
nsport.