Ll. Zamora et Jm. Calatayud, FLOW-INJECTION SPECTROPHOTOMETRIC DETERMINATION OF AMINO-ACIDS BASED ON AN IMMOBILIZED COPPER(II) ZINCON SYSTEM, Analytica chimica acta, 281(3), 1993, pp. 601-605
The flow-injection spectrophotometric determination of different amino
acids was carried out by reaction with copper(II) ions entrapped in a
polymeric material and filling a packed-bed reactor; the released cop
per(II), complexed with the amino acid, reacted with zincon in a basic
medium producing a blue colour that was monitored at 600 nm. The meth
od was applied to determine the contents of different amino acids in p
harmaceutical formulations. The calibration graph for glycine was line
ar over the range 0.5-20 mug ml-1 with a relative standard deviation o
f 0.8% (n = 6) at 10 mug ml-1 and a sample throughput of 108 h-1.