G. Knapp et al., Iodine determination in biological materials - Options for sample preparation and final determination, FRESEN J AN, 362(6), 1998, pp. 508-513
This paper describes the development of various new acid sample decompositi
on methods, as well as an extraction (leaching) method and compares them wi
th the "Schoniger Combustion" technique. The methods have been developed as
sample preparation techniques for iodine determination in biological mater
ials, especially in solid samples. ICP-MS (inductively coupled plasma mass
spectrometry) and a catalytic technique are employed and discussed fur the
final determination of iodine concentrations. Accuracy and reliability of t
he different analytical methods are shown in the examples of different CRMs
(certified reference materials) available for iodine. The re suits of an i
nterlaboratory comparison are specifically presented for the extraction (le
aching) method.