Many designs of unicompartmental knee replacement show early and mid-term f
ailure due to polyethylene wear. We studied the wear rate of congruent poly
ethylene meniscal bearings retrieved from failed Oxford unicompartmental kn
ee replacements,
We examined 16 bearings, 0.8 to 12.8 years after implantation, measuring th
eir thickness and comparing it with that of 14 unused bearings. The mean ra
te of penetration, which included the effects of wear at both upper and low
er surfaces, was 0.036 mm per year (maximum 0.08). Bearings as thin as 3.5
mm wore no faster than thicker models, but ten with evidence of impingement
had greater wear. The six bearings with no impingement showed a mean rate
of penetration of 0.01 mm psr year,
In unicompartmental knee replacement, careful implantation of fully congruo
us meniscal bearings can avoid failure due to polyethylene wear.