Experimental evaluation of nonlinearities of small-sized insertable gradient coils

Citation
D. Morvan et al., Experimental evaluation of nonlinearities of small-sized insertable gradient coils, MAGN RES IM, 16(10), 1998, pp. 1257-1263
Citations number
10
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging
Journal title
MAGNETIC RESONANCE IMAGING
ISSN journal
0730725X → ACNP
Volume
16
Issue
10
Year of publication
1998
Pages
1257 - 1263
Database
ISI
SICI code
0730-725X(199812)16:10<1257:EEONOS>2.0.ZU;2-D
Abstract
A phase imaging technique is proposed to map out and quantify gradient nonl inearities of small-sized insertable gradient coils, assuming the whole-bod y system gradients are highly linear in the domain of interest. The theory is developed and simple equations are derived to allow quantification. It i s applied to a 4-loop 18-cm diameter cylindrical gradient coil of optimal d esign. Experimental gradient nonlinearity maps are obtained for different f ields of view. Gradient non-linearities are quantified locally and in regio ns of interest, shelving close agreement with model data. (C) 1998 Elsevier Science Inc.