Calibration of the well-type germanium gamma-ray detector employing two gamma-ray spectra

Authors
Citation
M. Blaauw, Calibration of the well-type germanium gamma-ray detector employing two gamma-ray spectra, NUCL INST A, 419(1), 1998, pp. 146-153
Citations number
10
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
419
Issue
1
Year of publication
1998
Pages
146 - 153
Database
ISI
SICI code
0168-9002(199812)419:1<146:COTWGG>2.0.ZU;2-Y
Abstract
Since 1992, a method has been under development at IRI to characterise a we ll-type gamma-ray detector by determination of a peak efficiency and a peak -to-total efficiency curve through measurement of a radionuclide emitting c oincident photons, and to compute radionuclide specific counting efficienci es from those curves for "'real" peaks as well as for "sum" peaks. Recent i mprovements to this method have indicated that now, a well-type can be as e asy to calibrate and operate as a coaxial detector, once the software is in place. In this paper, the improved calibration method will be described an d examples of results given. (C) 1998 Elsevier Science B.V. All rights rese rved.