D. Nikezic et Kn. Yu, The influence of thoron and its progeny on radon measurements with CR39 detectors in diffusion chambers, NUCL INST A, 419(1), 1998, pp. 175-180
In this work partial and total sensitivities of the CR39 detector to thoron
and its progenies are determined by the Monte Carlo method. It has been sh
own that the total sensitivity to thoron is higher than that to radon, with
the difference in the sensitivities depending on the dimension and the sha
pe of the diffusion chamber used in measurements. The presence of thoron in
air causes an overestimation of the measured radon values. The overestimat
ion has been found to increase linearly with the ratio of thoron/radon conc
entrations and to depend on height and radius of the diffusion chamber. (C)
1998 Elsevier Science B.V. All rights reserved.