A method for analysing and comparing treatment planning system (TPS) data a
nd ferrous dosimeter gel measurements evaluated with MRI (FeMRI) was develo
ped, including image processing to final absorbed dose images. Measurements
were analysed according to this method and FeMRI data were thereby compare
d with the TPS-calculated dose distribution. For photons, differences betwe
en FeMRI- and TPS dose data were mainly within +/- 2%. Minor shortcomings f
ound in both the FeMRI system and the TPS are explained and discussed. For
electron beams, there was an overall good agreement. It was found that the
TPS underestimates the lateral scattering dose outside the primary beam, bu
t the reported dose difference corresponds to a small spatial deviation (le
ss than 2 mm). It is important to consider this single beam data comparison
when the method is extended to more complicated situations, for example wh
en using several beams.