Beryllium-selective membrane electrode based on benzo-9-crown-3

Citation
Mr. Ganjali et al., Beryllium-selective membrane electrode based on benzo-9-crown-3, ANALYT CHEM, 70(24), 1998, pp. 5259-5263
Citations number
49
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICAL CHEMISTRY
ISSN journal
00032700 → ACNP
Volume
70
Issue
24
Year of publication
1998
Pages
5259 - 5263
Database
ISI
SICI code
0003-2700(199812)70:24<5259:BMEBOB>2.0.ZU;2-C
Abstract
A PVC membrane electrode for beryllium(II) ions based on benzo-9-crown-3 as membrane carrier was prepared. The sensor exhibits a Nernstian response fo r Be2+ ions over a wide concentration range (4.0 x 10(-3)-2.5 x 10(-6) M) w ith a limit of detection of 1.0 x 10(-6) M (9.0 x 10(-3) ppm). It has a res ponse time of similar to 50 s and can be used for at least 4 months without any divergence in potential. The proposed electrode revealed very good sel ectivities for Be2+ over a wide variety of other cations including alkali, alkaline earth, transition, and heavy metal ions and could be used in a pH range of 2.0-6.0. It was successfully applied to the determination of beryl lium in a mineral sample.