S. Furlanetto et al., Set-up and validation of an adsorptive stripping voltammetric method for kynurenic acid determination in human urine, J PHARM B, 18(1-2), 1998, pp. 67-73
Validation of an adsorptive stripping voltammetric method for kynurenic aci
d determination in urine, was presented. The selection of appropriate valid
ation parameters, the design consideration for evaluation and the problem o
f endogenous metabolites were discussed. The considered fundamental criteri
a for assessing the reliability and overall performance of the method in th
e urine matrix were selectivity, linearity and range, limit of quantitation
, accuracy, precision and analyte stability. The intermediate precision was
also evaluated by means of a full factorial design. An HPLC method with fl
uorimetric detection was used as a reference method to assess the accuracy.
The analysis in urine required a pH control as pointed out by robustness t
esting and the found kynurenic acid concentration in daily urine ranged fro
m 5 to 40 mu M. (C) 1998 Elsevier Science B.V. All rights reserved.