Application of a video-optical beam imaging system for quality assurance of medical accelerators

Citation
Lj. Ma et al., Application of a video-optical beam imaging system for quality assurance of medical accelerators, PHYS MED BI, 43(12), 1998, pp. 3649-3659
Citations number
10
Categorie Soggetti
Multidisciplinary
Journal title
PHYSICS IN MEDICINE AND BIOLOGY
ISSN journal
00319155 → ACNP
Volume
43
Issue
12
Year of publication
1998
Pages
3649 - 3659
Database
ISI
SICI code
0031-9155(199812)43:12<3649:AOAVBI>2.0.ZU;2-3
Abstract
Method validation techniques were developed and experiments were carried ou t using a beam imaging system (BIS, Wellhofer Dosimetrie, Schwarzenbruck, G ermany) for routine quality assurance of medical accelerators. The routine quality assurance tasks include x-ray beam flatness and symmetry check, lig ht/radiation field congruence test, beam energy constancy for electrons and mechanical checks for couch and collimator rotations. Comparisons were mad e between the BIS application and conventional quality assurance methods th at use radiographic films or detector arrays. In this work, we have demonst rated efficiency and accuracy of the BIS to perform some of the routine qua lity assurance tasks for medical linear accelerators.