L. Cozzi et al., Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapy, RADIOTH ONC, 48(3), 1998, pp. 335-338
Citations number
10
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Onconogenesis & Cancer Research
The Hounsfield units into density conversion for a commercial treatment pla
nning system (TPS) is discussed. The impact of an average calibration table
provided by the manufacturer on computed doses was investigated when a cus
tomer calibration could not be included in the TPS. The maximum error deter
mined on the computed monitor units per Gy was found to be about 2%, being
below 1% on average. The relative importance of the parameters used for CT
image acquisition is also discussed. The applied voltage was the most relev
ant parameter leading to errors in the reconstructed Hounsfield numbers of
about 300 units for high densities. (C) 1998 Elsevier Science ireland Ltd.