Flow-through fluorescence immunosensor for atrazine determination

Citation
E. Turiel et al., Flow-through fluorescence immunosensor for atrazine determination, TALANTA, 47(5), 1998, pp. 1255-1261
Citations number
12
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
TALANTA
ISSN journal
00399140 → ACNP
Volume
47
Issue
5
Year of publication
1998
Pages
1255 - 1261
Database
ISI
SICI code
0039-9140(199812)47:5<1255:FFIFAD>2.0.ZU;2-#
Abstract
A new flow-through fluoroimmunosensor for atrazine determination based on t he use of protein A immobilized on controlled pore glass as immunoreactor i s reported. The support, placed in the optical path of the flow cell, allow s the 'in situ' quantification of atrazine by on-line antigen-antibody bind ing upon successive injections of both substances. The immunosensor has a d etection limit of 2.1 mu g l(-1), a sample speed of about 10 samples per ho ur, and provides high reproducibility both within-day (3.2% for 5 mu g l(-1 ) and 2.2% for 30 mu g l(-1)) and between days. The optimum working concent ration range was 2.1-50 mu g l(-1). Possible interferences of other triazin es like simazine, desethylatrazine (DEA) and desisopropylatrazine (DIA) wer e evaluated. Simazine and DIA were not cross-reactive; however, the cross-r eactivity for DEA was CR = 7.7%. The proposed immunosensor was successfully applied to the determination of atrazine in drinking water and citrus frui ts. (C) 1998 Elsevier Science B.V. All rights reserved.